Technical data
1Service
36 Agilent 1000A Series Oscilloscopes Service Guide
Channel 2
2mV/div ±6mV +0.96 to +1.04
5 mV/div ±15 mV +0.96 to +1.04
10 mV/div ±30 mV +0.97 to +1.03
20 mV/div ±60 mV +0.97 to +1.03
50 mV/div ±150 mV +0.97 to +1.03
100 mV/div ±300 mV +0.97 to +1.03
200 mV/div ±600 mV +0.97 to +1.03
500 mV/div ±1.5 V +0.97 to +1.03
1 V/div ±3.0 V +0.97 to +1.03
2 V/div ±6.0 V +0.97 to +1.03
5 V/div ±15.0 V +0.97 to +1.03
Channel 3
2mV/div ±6mV +0.96 to +1.04
5 mV/div ±15 mV +0.96 to +1.04
10 mV/div ±30 mV +0.97 to +1.03
20 mV/div ±60 mV +0.97 to +1.03
50 mV/div ±150 mV +0.97 to +1.03
100 mV/div ±300 mV +0.97 to +1.03
200 mV/div ±600 mV +0.97 to +1.03
500 mV/div ±1.5 V +0.97 to +1.03
1 V/div ±3.0 V +0.97 to +1.03
2 V/div ±6.0 V +0.97 to +1.03
5 V/div ±15.0 V +0.97 to +1.03
Table 13 DC Gain Test (continued)
Vertical
Sensitivity
Power
Supply
Setting
VDMM+ VDMM- VScope+ VScope- Calculated
DC Gain
Offset Gain Test
Limits