Technical data

Service 1
Agilent 1000A Series Oscilloscopes Service Guide 35
Performance Test Record
Table 12 Performance Test Information
Serial No. ______________________________________ Test by _____________________________
Test Interval ____________________________________ Work Order No. ______________________
Recommended Next Testing ________________________ Temperature ____________
Table 13 DC Gain Test
Vertical
Sensitivity
Power
Supply
Setting
VDMM+ VDMM- VScope+ VScope- Calculated
DC Gain
Offset Gain Test
Limits
Channel 1
2mV/div ±6mV +0.96 to +1.04
5 mV/div ±15 mV +0.96 to +1.04
10 mV/div ±30 mV +0.97 to +1.03
20 mV/div ±60 mV +0.97 to +1.03
50 mV/div ±150 mV +0.97 to +1.03
100 mV/div ±300 mV +0.97 to +1.03
200 mV/div ±600 mV +0.97 to +1.03
500 mV/div ±1.5 V +0.97 to +1.03
1 V/div ±3.0 V +0.97 to +1.03
2 V/div ±6.0 V +0.97 to +1.03
5 V/div ±15.0 V +0.97 to +1.03