Technical data

1Service
32 Agilent 1000A Series Oscilloscopes Service Guide
To test trigger sensitivity
This test verifies the trigger sensitivity. In this test, you apply a sine wave
to the oscilloscope at the bandwidth limits. You then decrease the
amplitude of the signal to the specified levels, and check to see if the
oscilloscope is still triggered.
Table 10 Trigger Sensitivity Specification
Trigger sensitivity Ch 1, 2, 3, 4 (DC
coupling):
5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to
full bandwidth
< 5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to
20 MHz
Table 11 Equipment Required for Trigger Sensitivity Test
Description Critical Specifications Recommended
Model/Part Numbers
Signal Generator 100 kHz to 1 GHz at 200 mVrms, 0.01 Hz
frequency resolution, jitter: < 2 ps
Agilent 8648A,
Agilent E4400B, or
Agilent N5181A.
BNC Cable 50Ω characteristic impedance, BNC (m)
connectors
Agilent 8120-1840,
Agilent 10503A
Type-N to BNC
Adapter
Type-N (m) to BNC (f) Agilent 1250-0780
50Ω BNC Feed
Through Adapter
50Ω BNC (f) to BNC (m) feed through
terminator
Agilent 0960-0301