Technical data

1Service
30 Agilent 1000A Series Oscilloscopes Service Guide
1 Set up the signal generator:
a Set the output to 10 MHz, approximately 1 Vpp sine wave.
2 Connect the output of the signal generator to oscilloscope channel 1
using the BNC cable.
Table 9 Equipment Required for Time Scale Accuracy Test
Description Critical Specifications Recommended
Model/Part Numbers
Signal Generator 100 kHz - 1 GHz, 0.01 Hz frequency resolution,
jitter: < 2 ps
Agilent N5181A,
Agilent E4400B, or
Agilent 8648A
BNC Cable 50Ω characteristic impedance, BNC (m)
connectors
Agilent 8120-1840,
Agilent 10503A
Type-N to BNC
Adapter
Type-N (m) to BNC (f) Agilent 1250-0780
50Ω BNC Feed
Through Adapter
50Ω BNC (f) to BNC (m) feed through
terminator
Agilent 0960-0301
Figure 9 Connecting Equipment for Time Scale Accuracy Test
Oscilloscope
Signal Generator
BNC cable
Type-N (m) to
BNC (f) adapter
50W feed
through