Technical data
Service 1
Agilent 1000A Series Oscilloscopes Service Guide 25
9 Press [Acquire].
10 Press the Acquisition softkey until “Average” appears.
11 Press Average and turn the entry knob until “8” appears.
12 Press [Measure].
13 Press the Voltage softkey.
14 Turn the entry knob to select the Vpp menu item.
15 Set the signal generator to a 1 MHz sine wave with a peak- to- peak
amplitude of about 6 divisions as it appears on the oscilloscope screen.
Figure 7 Averages Menu Item