Technical data
Service 1
Agilent 1000A Series Oscilloscopes Service Guide 23
1 Connect the equipment as shown in Figure 4.
2 Preset and calibrate the power meter according to the instructions
found in the power meter manual.
3 Set up the Power Meter to display measurements in units of Watts.
4 Press the [Default Setup] front panel key.
5 Press the [Auto-Scale] front panel key.
6 Set the channel 1 probe attenuation to 1X.
7 Set the channel 1 vertical scale to 200 mV/div.
Oscilloscope
Power Meter
E4418B
Power sensor
cable
Signal Generator
Power splitter
11667B
Type-N (m) to
SMA (f) adapter
Power sensor
8482A
SMA cable
SMA to BNC adapter
50W feed
through
Figure 4 Connecting Equipment for Maximum Frequency Check Test