Technical data

1Service
14 Agilent 1000A Series Oscilloscopes Service Guide
Before Testing Performance
Table 2 Equipment Required to Perform All Tests
Description Required for Test Critical Specifications Recommended
Model/Part Numbers
Digital Multimeter DC Gain Accuracy DC voltage measurement accuracy better
than ±0.1% of reading
Agilent 34401A
Power Supply DC Gain Accuracy 0 V to 35 V DC; 10 mV resolution Agilent E3633A or
Agilent E3634A
Signal Generator Analog Bandwidth
Time Scale Accuracy
Trigger Sensitivity
100 kHz to 1 GHz at 200 mVrms, 0.01 Hz
frequency resolution, jitter: < 2 ps
Agilent 8648A,
Agilent E4400B, or
Agilent N5181A.
Power Meter Analog Bandwidth Agilent E-series with power sensor
compatibility
Agilent E4418B
Power Sensor Analog Bandwidth 100 kHz to 1 GHz ±3% accuracy Agilent 8482A
Power Splitter Analog Bandwidth Outputs differ by < 0.15 dB Agilent 11667B
BNC Cable DC Gain Accuracy (qty 2)
Time Scale Accuracy
Trigger Sensitivity
50Ω characteristic impedance, BNC (m)
connectors
Agilent 8120-1840,
Agilent 10503A
BNC Tee Adapter DC Gain Accuracy BNC Tee (m)(f)(f) Agilent 1250-0781
BNC to Dual Banana
Adapter
DC Gain Accuracy (qty 2) BNC (f) to dual banana Agilent 1251-2277
SMA Cable Analog Bandwidth SMA (m) to SMA (m) 24 inch
50Ω BNC Feed
Through Adapter
Analog Bandwidth
Trigger Sensitivity
50Ω BNC (f) to BNC (m) feed through
terminator
Agilent 0960-0301
Type-N to SMA
Adapter
Analog Bandwidth Type-N (m) to SMA (f) Agilent 1250-1250
SMA to BNC Adapter Analog Bandwidth SMA (m) to BNC (m) Agilent 1250-0831
Type-N to BNC
Adapter