Specifications

74
1280C GB 20001204
Troubleshooting checklist
An interval beep accompanies the messages.
Error status Refer to Page
You try to feed a film but it is not
accepted.
The processor is out of opera-
tion.
77
Display messages ‘Developer level
low’ or ‘Fixer level low’.
The developer or fixer level is
too low.
79
Display messages ‘Water level low’. The water level is too low.
84
An accepted film is not deposited in
the exit tray.
Jammed film.
85
The processed film is still moist, or is
deformed or shows drying patterns.
Film quality problems.
87
Display messages ‘Check water
pump’.
Clean the water pump.
If that does not help, contact
your local service organization
.
68
‘Error 502 - See manual’ message.
The developer temperature is
too low.
88
‘Error 504 - See manual’ message. The fixer temperature is too low.
90
‘Error 514 - See manual’ message.
Clean the pumps. (DEV, FIX1,
FIX2)
If that does not help, contact
your local service organization.
68
‘Error 510 - See manual’ message.
Failure of the dryer heaters.
Procedure for skilled service
personnel only.
101
‘Error 501’, ‘Error 503’, ‘Error 510’,
‘Error 506’ and other ‘Error XXX’
messages.
Contact your local service organization.
Messages when accessories are connected.
92