Specifications
47
1280C GB 20001204
Setting the start-up replenishment function
This function assures an automatic developer and fixer replenishment every
time the processor is switched on. This setting should be activated if the proces-
sor is used on an irregular basis (only a few days per week).
The default setting is OFF.
Refer to the table for recommended start-up developer and fixer replenishment
settings, depending on the specific workload.
Related setting: Developer replenishment rate
more than 5 m² film/day 3 to 5 m² film/day less than 3 m² film/day
daily use irregular use daily use irregular use daily use irregular use
400 ml/m² 400 ml/m² 600 ml/m² 600 ml/m² 600 ml/m² 600 ml/m²
Related setting: Standby developer and fixer replenishment - jog cycle
more than 5 m² film/day 3 to 5 m² film/day less than 3 m² film/day
daily use irregular use daily use irregular use daily use irregular use
OFF OFF OFF OFF ON ON
Start-up replenishment function
more than 5 m² film/day 3 to 5 m² film/day less than 3 m² film/day
daily use irregular use daily use irregular use daily use irregular use
OFF ON OFF ON OFF ON
The start-up replenishment rate is not adjustable during film processing.
If you do not comply with the recommended replenishment rates, the
optimal processing stability is no longer guaranteed and sensitometric
variations might occur.