Service manual

Avantra 36/44 Service Manual
4-16 Maintenance
4.5 Service Program Codes
Code Function Switch Element 15% Key 85% Key
111 Main motor K 12 Start Stop
112 Main motor Speed adjust Increase Decrease
113
121 Circ. pumps K 11 Start Stop
122
123 Solenoid, water K 10 Open Close
131 Dev. repl. K 8 On Off
132 Fixer repl. K 9 On Off
133
211 Dev. heat LO Q 1 On Off
212 Fixer heat LO Q 3 On Off
213 Dev. heat HI Q 5 On Off
221 Dev. cooling (Optional) On Off
222
223 Fixer heat HI K 2 On Off
231 Dryer heat basic/fan K 7 On Off
232 Dryer heat 1/fan Q 4 + K 7 On Off
(50% key to
switch off K 7)
233 Dryer heat 2/fan Q 6 + K 7 On Off
(50%key to
switch off K 7)
311 Dev. repl. Pump pulses 10 pulses 1 pulse=100ml
312 Fixer repl. Pump pulses 10 pulses 1pulse= 100ml
313
321 Dev. repl. pump adjust K 8 (adj. 0–200%) +% –%
322 Fix. repl. pump adjust K 9 (adj. 0–200%) +% –%
323
Lamp Color
331 Warning signals Dev. temp. incorrect Red Blinking Off
Low level Red Blinking Off
Repl. empty Red Blinking Off
Wasteful Red Blinking Off
Input Lamp Color
332 Sensor read out Left input sensor set Film input Red
(left)
Right input sensor set Film input Red
(right)
Exit sensor Film input Red
(left & right)
333 Display test All dots shining Display Red
Table 4-1 Service program codes.