Service manual

Section IV: On-line Processor
Maintenance 4-13
c. To adjust the developer, use program 321. To select the program,
press:
WAT
FIX
DEV
Note: DEV = 1
FIX = 2
WAT = 3
The display shows a preset value from 0 to 200%. Average values
range from 95% to 105%.
d. Press the 15% key to increase the percentage value and the 85% key to
decrease the percentage value. Adjust the value so that the amount of
replenisher released will be 100 ml.
e. To adjust the fixer, use program 322. Press:
WAT
FIX
FIX
Note: DEV = 1
FIX = 2
WAT = 3
Repeat Step 2d from this procedure.
NOTE: After a percent change has been made, check the output by
entering program 311 for developer or 312 for fix. Pressing
the 85% key releases one pulse that should equal 100 ml. If
it does not, repeat step 2d.
NOTE: The adjustment range is 0-200%. If you reach a value of
200% and are not getting 100 ml per pulse, it may be neces-
sary to increase the pump stroke. Loosen the nut on the
bottom of the pump and move the pump coil assembly
toward the spring. Retighten.
f. Press the on/off key to exit the service program.