Service manual

Section IV: On-line Processor
Maintenance 4-7
Figure 4-2 New Control Panel with icon key pad
(Descriptions of keys on next page)
OLP Control Panel Icons (for Figure 4-2
)
1Low Dev. or Fix. 8Temp. 15Enter 22Sound
2Display 9Stdby Repl. 16Man. Start 23Water
3Low Repl. 10Repl. 17Man. Repl. 24Film Feed Right
4Waste Full 11Up 18Dryer Saver 25Exposure %
5Fix. 12Preset 19Auto 26Dev. Temp. Incorr.
6Dev. 13Down 20Panel Light 27P1-P9
7Time 14On/Off 21Dryer 28Film Feed Left
4.3.1 LUTH Motor Control Board (LMB) 24V DC (P/N
5202Y000100)
The LMB 24V performs the following functions:
Amplifies the motor preset value to the corresponding voltage for driving
the main motor
Receiving and converting /R pulse preset values (optical link)
Receiving and converting tacho feedback pulses
NOTE:
The following procedure is for OLPs with serial
#3613DD300659 and above. See Section 4.3.2 for calibrat-
ing the 180V LMB.
The following procedure calibrates the LMB (refer to Figure 4-3). A shorter remote
control panel cable is taped to the top of the electronics box to make the procedure
easier.
To Reposition the Control Panel:
1. Remove all power to the processor.
Dev
Fix
H
2
O
Dev
85%50%15%
1
2 3 4 5
6
7
8 9
10
11 12 13
141516
1718
19
2021222324
25
2627
28
P1-P9