Service manual

Section IV: On-line Processor
Maintenance 4-3
Checking Temperatures
Below are some examples of the correlation between temperature and resistance:
21
°
F (-6.1
°
C) = 540.4 ohm
30
°
F (-1.1
°
C) = 557.8 ohm
32
°
F (0.0
°
C) = 561.6 ohm
34
°
F (1.1
°
C) = 566.5 ohm
35
°
F (1.7
°
C) = 569.3 ohm
38
°
F (3.3
°
C) = 573.2 ohm
40
°
F (4.4
°
C) = 577.0 ohm
42
°
F (5.6
°
C) = 580.9 ohm
45
°
F (7.2
°
C) = 586.6 ohm
50
°
F (10.0
°
C) = 596.3 ohm
55
°
F (12.8
°
C) = 605.9 ohm
60
°
F (15.6
°
C) = 615.5 ohm
NOTE:
If temperatures in the bath are incorrect, no adjustment is
possible. The only corrective action available is to replace
the defective temperature sensor.
4.2.2 Checking Operation of the LCB
WARNING:
All AC power must be off prior to removing the X400
plug.
1. Check the power supply circuits:
a. Disconnect the X400 plug Agfa Terminal Board (ATB).
i. Measure the voltage between pins 4 and 5. The nominal value is
18 VAC.
ii. Measure the voltage between pins 4 and 3. The nominal value is
12 VDC.
iii. Reconnect the X400 plug.