Service manual

Section IV: On-line Processor
Installation 1-15
Figure 1-11 Assembling the legs, brace and leveler feet.
–Support leg.
–Two hex screws.
–Stabilizer brace.
–Rear leveler feet (block legs and attach these last).
1.5.2 Connecting the Water Supply and Drainage
CAUTION: To ensure sufficient rinsing of the media after developing
and fixing, the processor must be connected to temperate wash
water. The water temperature must not exceed 110
°
F (43
°C), and the
water pressure must be 0.3 bar to 10 bar. Filtration rate of the filter
cartridge should be at least 25 microns.
CAUTION: The temperature difference between the fixer and the
water in the wash section must not exceed 27°F (15°C). Exceeding
this temperature may cause stress of material and bending of the
partition between the fixer and wash sections.