Installation guide
User migrations and configurations are performed on the Application Server (as described in the
AS300 Preparing to Upgrade Guide—IMPAX 4.5, 5.2, 5.3, or WEB1000 to IMPAX 6.5.3) and worklist
and report data can be migrated from the training server (described in Migrating data from the
training server (refer to page 32)).
After all these migrations are complete, reconfigure the Application Server to connect to the
production database instead of the training server. Several steps are required:
1. Disable the connection to the temporary IMPAX 6.5.3 database, to the IMPAX 5.2 or 5.3
database and, if applicable, to the WEB1000 Server.
2. Connect to the production IMPAX 6.5.3 database.
3. Import the portable password file generated from the production 6.5.3 Database Server.
4. Set the password and account lockout policies.
5. Connect to a non-queryable RIS and remove an IP address from the IP filtering list.
To connect to another type of RIS (local or remote Agfa RIS or a queryable RIS), refer to
instructions in the IMPAX 6.5.3 Application Server Knowledge Base.
6. Perform other Application Server configurations that could not be completed during the
preparing to upgrade period, such as managing web services, setting up Healthcheck, and
managing SSL certificates.
Details on each of these steps follow.
Disabling SQL connections
(Topic number: 6802)
1.
The Application Server is currently configured to communicate with the IMPAX 5.2 or 5.3 SQL
Server database. Disable the connections to those servers, and to the WEB1000 Server station, if you
have set that up.
To disable SQL connections
1.
On the Application Server, open the Windows Administrative Tools and select Data Sources
(ODBC).
2.
Switch to the System DSN tab.
3. Select the name of the IMPAX 5.2 or 5.3 database.
4.
Click Remove. Confirm the removal.
5. If a connection to the WEB1000 Server was set up, select the name of that database and click
Remove. Confirm the removal.
6.
Click OK.
Connecting to the IMPAX 6.5.3 SQL Server database
(Topic number: 6811)
2.
You must now configure the Application Server for the migrated IMPAX 6.5.3 database.
AS300 Upgrade Guide—IMPAX 5.2 or 5.3 to IMPAX 6.5.3 40
Agfa Company Confidential—Subject to Non-Disclosure Obligation