Installation guide

Connectivity Manager outbound message queues must be configured with the new server settings
before messages are added to the queues. Consult a Connectivity Manager integrator to create a
device for the destination IMPAX server. Report updates can be sent to only one IMPAX server,
after all reports have been copied to that server.
Stop transmitting data to IMPAX
(Topic number: 7617)
7.
Allow remaining SEND jobs to continue until they have finished, then stop any more studies from
being transmitted in the IMPAX system.
To stop transmitting data to IMPAX
1.
Open the Windows Administrative Tools and select Services.
2.
Right-click the DICOM Service Class Provider service and select Properties.
3.
To change the Service status, click Stop.
4.
From the Startup type list, select Disabled.
5.
To close the Properties dialog, click OK.
6.
Launch the IMPAX Service Tools and log in as user service.
7.
On the Daily tab, select Job Manager.
8.
Monitor each Transmit queue and wait for all outgoing jobs to finish.
You cannot delete jobs in progress.
9.
Select each Transmit queue and click Halt Queue.
10.
To confirm that you want to halt the queue, click Yes.
Redirecting studies to the training server
(Topic number: 10235)
8.
Configure modalities to redirect studies to the training server system, so that they remain accessible
to the IMPAX 6.5.3 Clients while the migration continues.
The details of how to redirect studies are modality-specific and are not covered in this guide.
AS300 Upgrade Guide—IMPAX 5.2 or 5.3 to IMPAX 6.5.3 15
Agfa Company Confidential—Subject to Non-Disclosure Obligation