Installation guide
The temporary DAA cluster will need an Application Server, so move an existing one to this cluster,
install a new server with the Application Server software from the same version of IMPAX as the
production cluster, or if supported, install the Application Server software on the temporary DAA
Database Server. For instructions, refer to the appropriate installation guide: IMPAX 6.2 Application
Server Installation and Upgrade Guide, IMPAX 6.3 Application Server Installation and Upgrade Guide,
or IMPAX 6.4 Application Server Installation and Configuration Guide.
Emptying Connectivity Manager queues
(Topic number: 113307)
4.
Before shutting down IMPAX to upgrade the system, empty all DM Out or impax_report_server
queues. Consult Connectivity Manager service personnel to discuss queues that have error
transactions.
To empty Connectivity Manager queues
1.
In Connectivity Manager, open Service Tools and click Queue Manager.
2. Select any device with either pending or error transactions and empty the queues.
3. Retry recent messages and delete older messages, since newer transactions may have updated
patient, study, and report data after these transactions entered an error state.
Stopping Connectivity Manager interfaces
(Topic number: 113766)
5.
During the IMPAX upgrade, you can prevent the loss of clinical patient updates from hospital
information systems by stopping data bound for the Connectivity Manager, or by stopping the
Connectivity Manager’s outbound queues. The preferred method is to stop inbound interfaces,
which prevents the Connectivity Manager from receiving incoming messages.
Coordinate with hospital information system personnel to confirm that they are capable of holding
messages in queues. If the information system queues can be stopped, also stop the Connectivity
Manager’s inbound interfaces.
To stop Connectivity Manager interfaces
1.
In Connectivity Manager, open Service Tools.
The Device Manager displays a list of devices and interfaces and their status.
2.
To re-sort and group all device classes, click Class.
3. Scroll down to view CMSI and HL7 class devices.
4. Note which HL7 In and CMSI In interfaces are started. These interfaces must be restarted after
the IMPAX upgrade.
5. Select the checkbox beside each of the started inbound interfaces.
6.
Click Stop.
AS300 Upgrade and DAA Consolidation Guide—IMPAX 6.2 or 6.3 to IMPAX 6.5.3 17