Installation guide

where remote_port_value is the TCP port where the Application Server listens for signals.
a. By default, the port number is 6000. To confirm the port number, navigate to the folder
where study-status-relay.bat is installed (for example, C:\Program Files\Agfa\Impax
Business Services\study-status-signal-relay) and open the Study.Status.Relay.exe.config
file in a text editor.
b. The port number value can be found under <StudyStatusRelayConfiguration> and
between the <SignalListenerPort> and </SignalListenerPort> tags.
4. Optionally, change the default values of the external_port, rad_user, or trainee_user keys.
DescriptionINI_VALUEINI_KEYINI_SECTION
The TCP port to listen for
signals coming in from the
A valid port to listen for
incoming signals. Default
value is 6000.
external_portsignal-relay
remote cluster (whereas
remote_port is used for
forwarding signals).
If you change this value, you
must also configure this user in
The user name used when
changing the study status as
rad_usersignal-relay
the IMPAX 5.2 or 5.3 Service
Tools.
a radiologist. The default
value is signal-relay.
If you change this value, you
must also configure this user in
The user name used when
changing the study status as
trainee_usersignal-relay
the IMPAX 5.2 or 5.3 Service
Tools.
a trainee radiologist. The
default value is sig-relay-train.
Copying the 6.5.3 Cross-Cluster Dictation Interlock components
(Topic number: 59189)
4.
Perform the following task on the IMPAX 6.5.3 Application Server.
To copy the 6.5.3 Cross-Cluster Dictation Interlock components
1. Copy the study-status-signal-relay folder from the Cross-Cluster Dictation Interlock zip in the
Tools subdirectory to an appropriate folder on the 6.5.3 Application Server, such as to
C:\Program Files\Agfa\Impax Business Services.
The service will run from anywhere, but you should place it in a folder that is unlikely to be
deleted.
2. Open a command prompt and change to the directory containing the copied files.
3.
Change to the study-status-signal-relay subdirectory.
4. Type
import-study-status-relay.bat
5.
In the Apply Study Status Relay ADAM Schema dialog, click OK.
AS300 Upgrade and DAA Consolidation Guide—IMPAX 5.2 or 5.3 to IMPAX 6.5.3 55