Installation guide
5.
Open the Windows Administrative Tools and select Services.
6. Start the service.
The install_signal_relay.bat file installs the mvf_signal_relay.exe as a Windows service, inserts default
values into the map_ini, and creates a radiologist and a trainee radiologist user. These users are used
by the service to change status when appropriate. The radiologist user is called signal-relay and the
trainee radiologist is called sig-relay-train.
Updating map_ini values for Cross-Cluster Dictation Interlock
(Topic number: 48091)
For the connection between the 5.2 or 5.3 cluster and the IMPAX 6.5.3 cluster to work, some of the
default values added to the 5.2 or 5.3 IMPAX MAP_INI file must be changed.
To update map_ini values for Cross-Cluster Dictation Interlock
1. Launch CLUI.
2. To update the remote hostname, type
update map_ini set ini_value='remote_host_name' where ini_section='signal-relay' and
ini_key='remote_host'
where remote_host_name is the name of the IMPAX 6.5.3 Application Server where the service
will run.
3. To update the remote port value, type
update map_ini set ini_value='remote_port_value' where ini_section='signal-relay' and
ini_key='remote_port'
where remote_port_value is the TCP port where the Application Server listens for signals.
a. By default, the port number is 6000. To confirm the port number, navigate to the folder
where study-status-relay.bat is installed (for example, C:\Program Files\Agfa\Impax
Business Services\study-status-signal-relay) and open the Study.Status.Relay.exe.config
file in a text editor.
b. The port number value can be found under <StudyStatusRelayConfiguration> and
between the <SignalListenerPort> and </SignalListenerPort> tags.
4. Optionally, change the default values of the external_port, rad_user, or trainee_user keys.
DescriptionINI_VALUEINI_KEYINI_SECTION
The TCP port to listen for
signals coming in from the
A valid port to listen for
incoming signals. Default
value is 6000.
external_portsignal-relay
remote cluster (whereas
remote_port is used for
forwarding signals).
If you change this value, you
must also configure this user in
The user name used when
changing the study status as
rad_usersignal-relay
AS300 Preparing to Upgrade Guide—IMPAX 4.5, 5.2, 5.3,or WEB1000 to IMPAX 6.5.3 66