User Manual Part 2

119
2852 A EN 20070205
DRYSTAR AXYS
5 Record these values on Chart 3 (‘Drystar AXYS Artifacts and Spatial Resolution
Control Chart’). Refer to “Charts for mammography QC (optional)” on
page 155.
WARNING: In case of significant artifacts, insufficient spatial
resolution or failure of any other recommended QC tests, the
cause of the problem must be identified, and corrective action
must be taken before the Drystar AXYS can be used for any further
clinical imaging.
WARNING: Refer to “Preventive maintenance schedule” on
page 121 and “Maintaining image quality and resolving image
quality problems” on page 258 (Reference manual), or call your
local Agfa service organization.