User Manual Part 2
108
2852 A EN 20070205
DRYSTAR AXYS
Establishing the daily operating reference density
levels for mammography application
(DT 2 Mammo) (optional)
This procedure enables you to establish the base line values for:
Q Low density
Q Mid density
Q High density
Procedure
1 Press the Key-operator key to enter the Key-operator mode.
2 Press the down key seven times, followed by the confirm key to select ‘Quality
Control’.
A confirmation screen appears:
WARNING: The densitometer of the Drystar AXYS is calibrated at
installation. Authorized service personnel should recalibrate the
internal densitometer annually or after major service or repair.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation