User's Manual

95
2852 A EN 20070205
DRYSTAR AXYS
Establishing the image geometry reference values
for general radiography applications (DT 2 B and
DT 2 C)
Procedure
1 Print the QC test image or use the previously printed test image.
For an example, see “QC test image for general radiography applications (DT 2 B and
DT2C) on page 90.
2 To determine the reference values for geometry, measure the distances
A and B of the geometric square on the test image.
3 Record these values as reference dimensions Aref and Bref on Chart 4
(‘Drystar AXYS Geometric Consistency Control Chart’). Refer to “Quality Con-
trol Charts” on page 149.
These charts will be used for the annual quality test. For more information, refer to
“Performing the Annual QC tests for general radiography applications (DT 2 B and
DT2C) on page 103.
4 Save this film for future reference.
WARNING: Make sure to measure distance A from the left edge of
the left line to the right edge of the right line and distance B from
the upper edge of the upper line to the lower edge of the lower
line.
WARNING: We strongly recommend using a 30 cm (12-inch)
machinist scale with 0.5 mm divisions (1/64 inch).