2831A EN 20041201Advanced operation (key-operator mode)
DRYSTAR 5302 USER MANUAL
After the image is printed, the system will display the optical density values:
5 Record the density values on the Drystar 5302 Charts 2A and 2B (Drystar 5302
Daily Density Control Chart’). Also record the date and time of the test on the
charts and on the QC test images. Refer to ‘Quality Control Charts’ on page 75.
6 Press the ok key. The following screen is displayed:
7 Press the ok key to return to the main menu.
In case the measure results are not within the aim values, the reason for
the unacceptable density variations must be identified and resolved before
any further clinical films can be processed. This may include repeating the
film calibration procedure.
For possible causes of non-compliance and the respective actions, refer to
‘Preventive maintenance schedule’ on page 59 and ‘Maintaining image quality
and resolving image quality problems’ on page 17 3 of the Drystar 5302
Reference manual.
QC
Internal Density readings
Low density :0.19
Mid density :1.25
High density: 1.78
(copy on control chart)
ok
Key-operator
QC
QC
Proceed with the QC proce-
dure as stated in the User man-
ual
ok
Key-operator
QC