2831A EN 20041201Advanced operation (key-operator mode)
DRYSTAR 5302 USER MANUAL
The displayed values represent the following steps on the test film:
Low density: the density of the Low density step.
Target: 0.4.
Mid density: the density value of the Mid density step.
Target: 1.2.
High density: the density value of the High density step.
Target: 2.0.
5 Record the density levels on the Drystar 5302 Chart 1 (‘Determination of the
operating levels’). Refer to ‘Quality Control Charts’ on page 75.
6 Press the ok key. The following screen is displayed:
If the mid density value does not meet or exceeds the recommended
values, the reason must be found and the problem solved before any
further clinical films can be printed.
Refer to ‘Preventive maintenance schedule on page 59 and Maintaining image
quality and resolving image quality problems’ on page 173 of the Drystar 5302
Reference manual, or call your local Agfa service organization.
QC
Proceed with the QC proce-
dure as stated in the User man-
ual
ok
Key-operator
QC