2831A EN 20041201 Advanced operation (key-operator mode)
DRYSTAR 5302 USER MANUAL
Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
Low density
Mid density
High density
To establish the daily operating levels, proceed as follows:
1 Press the Key-operator key to enter the Key-operator mode.
2 Press the down key seven times, followed by the ok key to select ‘QC’.
The ‘QC’ screen appears:
3 Press the ok key to continue.
The Drystar 5302 will automatically print the QC Test image.
4 After the image is printed, the system will display the optical density values:
The densitometer of the Drystar 5302 is calibrated at installation.
Authorized service personnel should recalibrate the densitometer
annually or after major service or repair.
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore configuration
6 Calibration
7 Installation
8 QC
quit
ok
select
Key-operator
main menu
QC
Start printing test image
to perform:
daily / weekly
and yearly control
quit
ok
Key-operator
QC
QC
Internal Density readings
Low density :0.26
Mid density :1.35
High density :1.89
(copy on control chart)
ok
Key-operator
QC