User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
99
2920B EN 20031206
5 Press the up/down arrow keys to select the desired save set file, followed by
the Confirm key.
The following confirmation screen is displayed:
6 Press the Confirm key (YES) to proceed with the selected file, or the Escape
key (NO) to quit the procedure without any changes.
The printer restores the configuration from the external CF-card:
When the copying of the save set is finished, the printer will ask you to remove the
CF-card:
7 Remove the CF card.
" Refer to ‘Removing a CF-card’ on page 22.
As soon as the CF-card has been removed, the following screen is displayed (when
necessary):
Press the Confirm key to reset the printer. After the reset, the new settings will be
active.
Restore
from CF-card
<Dataset def v1.00>
YES (v) NO (x)
RC
Copying
A:/<path><filename>
to
C:/<path><filename>
RC
Installation
finished
Please remove card
RC
New settings will
be active after
reset
RC