User's Manual Part 2

Table Of Contents
90
2920B EN 20031206
$ If you select “CubicBell”, the following screen will appear, allowing you to
adjust the smoothing factor:
Enter the smoothing factor value by means of the arrow keys. Refer to ‘Data entry
on page 27. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
Changing the maximum density (MaxD)
1
Perform steps 1 to 3 of the ‘Changing Image quality settings for a modality
procedure on page 87 to select the modality and choose ‘Select other’.
The ‘Select for Modality’ screen is displayed:
2 In the ‘Select for Modality’ screen, press the Down key twice, followed by the
Confirm key to select ‘MaxD’.
The ‘MaxD (maximum density)’ screen is displayed:
All maximum densities are selectable starting from 2.80 to 3.40 in steps of 0.05
units. The activated setting is marked with a ‘*’.
"
The maximum density will be limited to the maximum density of the media.
3 Press the up/down key to select the desired maximum density setting for this
modality.
4 Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
CS
Enter smoothing
parameter:
advised value = 0.35
range [0.1-10]
enter absolute value
y.yy
CS
Select for
Modality16char:
1 Look-up table
2 Interpolation
3MaxD
CS
Select MaxD for
Modality16char:
2.80
2.85
...
*3.10
3.15
...
3.40