User's Manual Part 2

Table Of Contents
87
2920B EN 20031206
Changing Image quality settings for a modality
1
Perform steps 1 and 2 of the ‘Accessing the Change settings menu’
procedure on page 59.
2 On the ‘Change settings’ screen, press the Down key three times, followed by
the Confirm key to select ‘Image quality’.
The following screen will appear:
3 Press the Up/Down key, followed by the Confirm key to select the modality for
which you want to change the Image quality settings.
" If the settings for the selected modality were set via a remote computer,
you will not be able to change the image quality information for this
modality manually. In this case the following screen will appear:
If the image quality parameters need to be changed for this modality, please use the
DICOM Configuration tool, in the specialist mode on the remote PC. Alternatively,
you can remove this modality (refer to ‘Removing a modality on page 79) and add a
new modality with the same name. For this ‘new’ modality new parameters can be
assigned via the local keypad (refer to ‘Adding a modality on page 80).
" If a file with default settings (‘Profile default’) for the selected modality
exists, the following screen is displayed:
" If you want to use these profile default settings for the current modality,
press the Confirm key. The Drystar 5300 will be loaded with these
settings and return to the ‘Image quality’ main menu.
CS
1 General setting
2 Def. densitometer
3 Modality16char
4 Modality16char
5 etc...
CS
Modality16char
Parameters can
only be changed by
DICOM configurator
CS
Default for
Modality16char:
1 Profile default
2 Select other