User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
85
2920B EN 20031206
Changing general image quality settings
1
Perform steps 1 and 2 of the ‘Accessing the Change settings menu’
procedure on page 59.
2 On the ‘Change settings’ screen, press the Down key three times, followed by
the Confirm key to select ‘Image quality’.
The following screen will appear:
3 Press the Confirm key to select ‘General setting’.
The ‘General setting’ screen is displayed:
This setting is by default set to ON and will apply to every modality. The currently
active setting is indicated with a ‘*’.
" For more information on film calibration settings: refer to ‘Understanding
the calibration policy’ on page 102.
4 Press the up/down key to select the film calibration default.
When the setting is ON, the Drystar 5300 will ask you to enter the calibration
frequency:
Enter the calibration frequency by means of the up/down key. Refer to ‘Data entry’
on page 27.
5 Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
CS
1 General setting
2 Def. densitometer
3 Modality16char
4 Modality16char
5 etc...
CS
Select default
film calibration
*ON
OFF
CS
Select Frequency
Every 01 film packs