User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
60
2920B EN 20031206
Each time the settings have been changed and confirmed, an automatic backup of
the old configuration is made on the internal CF-card. You will also be asked to
create a backup on an external CF-card.
After selecting YES, the ‘Update backup on CF-card’ screen is displayed:
If you want to save the configuration on the backup CF-card, press the Confirm key
(YES). The printer will now enter the ‘Save configuration procedure’ (see page 95).
If you only want to save the configuration on the internal CF-card, press the Escape
key (NO).
Next, the following screen is displayed (when necessary):
Press the Confirm key to reset the printer. After the reset, the new settings will be
active.
Update backup on
CF-card
YES (v) NO (x)
CS
New settings will
be active after
reset