User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
196
2920B EN 20031206
Drystar DT 2B
Legend:
(1) = X-RITE 310
(2) = GRETAG D200 II
(3) = MACBETH TR924
(4) = X-RITE 341
(5) = X-RITE 331
(6) = X-RITE 301
dens. (1) (2) (3) (4) (5) (6)
0% 0.25 0.18 0.23 0.21 0.22 0.20
10% 0.55 0.46 0.51 0.49 0.50 0.48
20% 0.84 0.75 0.80 0.78 0.79 0.77
30% 1.14 1.03 1.08 1.06 1.08 1.05
40% 1.43 1.32 1.37 1.36 1.37 1.34
50% 1.73 1.60 1.65 1.64 1.65 1.62
60% 2.02 1.89 1.94 1.93 1.94 1.91
70% 2.32 2.18 2.22 2.21 2.22 2.19
80% 2.61 2.47 2.51 2.51 2.52 2.48
90% 2.91 2.75 2.79 2.79 2.80 2.75
100% 3.20 3.04 3.08 3.08 3.09 3.04