User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
119
2920B EN 20031206
4 Press the Confirm key (YES) to start the QC procedure or the Escape key
(NO) to quit.
The Drystar 5300 will automatically print the QC test image.
After the QC test image is printed, the Drystar 5300 will display the optical density
values:
5 Report the density levels on the Drystar 5300 Control Charts 2A and 2B
(Drystar 5300 ‘Daily Density Control Chart’). Also record the date and time of
the test on the charts and on the QC test image. Refer to Appendix C ‘Quality
Control Charts’ on page 199.
6 Press the Confirm key to return to the main menu.
In case the measure results are not within the aim values, the reason for
the unacceptable density variations must be identified and resolved
before any further clinical films can be processed. This may include
repeating the film calibration procedure.
For possible causes of non-compliance and the respective actions, refer
to ‘Maintaining optimal image quality’ on page 101 and ‘Preventive
maintenance’ on page 161.
QC procedure
in progress
Please wait
62%
QC
QC readings:
Base+Fog x.xx
Mid Density x.xx
DD x.xx MaxD x.xx
(densitometer=xxxxx)
QC