User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
114
2920B EN 20031206
Verifying Acceptable Spatial Resolution and Artifact Levels
To establish the image geometry reference values, proceed as follows:
1 Print the 'QC ' test image or use the previously printed QC test image used to
establish the daily operating density levels.
You should obtain an image looking like this:
Drystar 5300 Quality Control Test Image
Date : __________Initials : _____
Sharpness Test Geometry Test
Density test
Dimension A
Dimension B
Each cluster
has 3 groups
of dots of
different sizes
Each group has 5 dots
3 Resolution clusters
MaxD (100%):
DD (Density Difference):
(High-Low)
Mid Density:
Base + Fog: