User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
111
2920B EN 20031206
Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
• Base+Fog density,
• Mid density,
• DD (Density Difference),
• MaxD (maximum density).
To establish the daily operating density levels, proceed as follows:
1 Press the Key-operator key to enter the key-operator mode.
2 Press the Down key seven times, followed by the Confirm key to select
‘Quality control’.
A confirmation screen appears:
3 Press the Confirm key (YES) to start the QC procedure or the Escape key
(NO) to quit.
The Drystar 5300 will automatically print the QC test image.
By default, the internal densitometer of the Drystar 5300 will be used. The
internal densitometer of the Drystar 5300 is calibrated at installation.
Authorized service personnel should recalibrate the internal
densitometer annually or after major service or repair.
Do not change the ‘default’ internal densitometer during the entire QC
procedure!
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore config.
6 Calibration
7 Service actions
8 Quality control
9 Installation
KO
Start QC
Procedure
YES (v) NO (x)
QC
QC procedure
in progress
Please wait
62%
QC