User's Manual Part 2
Table Of Contents
- Table of contents
- Introducing the Drystar5300
- Basic operation (Operator mode)
- Advanced operation (Key-operator mode)
- Overview of key-operator functions
- Viewing printer information
- Accessing the Show settings menu
- Viewing general information
- Viewing input tray information
- Viewing network (DICOM) information
- Viewing general image quality information
- Viewing default densitometer information
- Viewing image quality information of a modality
- Viewing installed data-sets information
- Viewing logical printers information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Maintaining optimal image quality
- Service actions
- Quality control procedure
- Installation operations
- Controlling the Drystar5300 via a remote PC (with browser)
- System description
- Preventive maintenance
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Index
106
2920B EN 20031206
6 Check the printed test film.
The test film should be similar to the image below.
Density response of Drystar media
Different densitometers can give different results for measuring the same
density area. This is due to spectral differences between the densitometers.
The density values used in this manual are based on the Macbeth TR924
densitometer. Values measured with another densitometer can vary from
these values.
Non-calibrated or badly calibrated densitometers can result in even larger
differences.
To give the user an idea of the differences given by the most frequently used
densitometers, you can find the measurements for an SMPTE pattern on the
Drystar DT 2 B and Drystar DT 2 C films in Appendix B.
The calibration film must be free of dust particles or any other artifacts. If
this is not the case, clean the print head (refer to ‘Print head cleaning’ on
page 107) and restart the calibration procedure.
In order to be usable this film should be absolutely dust free withinthe marked region!
FilmCalib53_14x17.BMP (Version2)