User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
93
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Changing general image quality settings - Default densitometer
1 Perform steps 1 and 2 of the ‘Accessing the Change settings menu’ procedure on
page 70.
2 On the ‘Change setting’ screen, press the Down key four times, followed by the
Confirm key to select ‘Image quality’.
The following screen will appear:
3 Press the Confirm key to select ‘General’.
The ‘General settings’ screen will appear:
4 Press the Down key once, followed by the Confirm key to select ‘Default
densitometer’.
The ‘Default densitometer’ screen is displayed:
The current default densitometer is marked with an *.
5 Press the Up/Down arrow keys to select the default densitometer.
6 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
All shown densities will vary slightly according to the densitometer selection. Refer
to ‘Drystar media density response data’ on page 237.
2 <Modality name>
3 <Modality name>
4 <Modality name>
5 <Modality name>
6 <Modality name>
7 <Modality name>
* Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
X-Rite 331
X-Rite 341