User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
88
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
The ‘Enter modality daily used name’ screen is displayed:
6 Enter the modality daily used name (nickname) by means of the arrow keys.
Refer to ‘Data entry’ on page 33. Make sure not to enter more than 8 characters.
7 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
The ‘Select modality brand’ screen is displayed:
8 Press the Up/Down arrow keys to select the appropriate modality brand name.
• In case you have selected ‘Other’, the Modality AE-title will be linked with the
default settings. Proceed with step 10.
9 Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
The ‘Select modality type’ screen is displayed:
10 The ‘select modality AE-title’ screen is displayed:
A blank daily used modality name is not allowed.
2 Philips
3 Fuji
4 Siemens
1 MXVIEW
2 MX8000
2 <Modality name>
3 <Modality name>
4 <Modality name>