User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
66
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Viewing general Image quality information
1 Perform steps 1 and 2 of the ‘Accessing the Show settings menu’ procedure, on
page 61.
2 On the Show settings menu, press the Down key four times, followed by the
Confirm key to select ‘Image quality’.
The following screen will appear:
3 Press the Confirm key to select ‘General’.
The general Image quality settings are displayed:
The following info is displayed:
• Film calibration setting: OFF/ON with the number of film packs
determining the calibration interval (if ON).
• Dmax upper tray: the maximum density for the upper input tray resulting
from the automatic film calibration.
• Dmax lower tray: the maximum density for the lower input tray resulting
from the automatic film calibration.
• TPH cleaning: the number of prints after the last thermal head cleaning,
and the number of prints to go before a new cleaning session is necessary.
• TPH profile: the number of prints after the last thermal head profile
calibration and the number of prints to go before a new thermal head
profile calibration is necessary.
• Def. densitometer: shows the default densitometer that is used for the
calibrations.
4 Press the Confirm key to return to the Image quality menu.
5 In the Image quality menu, press the Escape key to return to the Show settings
menu.
2 <Modality name>
3 <Modality name>
4 <Modality name>
5 <Modality name>
6 <Modality name>
7 <Modality name>
OFF
3.12
3.11
OFF
OFF
Macbet