User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
222
2900H EN 20071108Troubleshooting
DRYSTAR 5500/5503 REFERENCE MANUAL
The following flowchart will assist you in determining the correct operation
in order to maintain the optimal image quality from your printer:
Image
Quality OK?
NO
YES
Persistent
White Dots or
Lines?
Start
Call Service
Done
Clean Dust
Rollers
YES
NO
Clean
Print Head
Image
Quality
OK?
Image
Quality OK?
NO
YES
Done
YES
Done
Perform Film Calibration
Perform Print Head
Profile Calibration
Perform Film Calibration
NO