User Manual

Table Of Contents
221
2900H EN 20071108 Troubleshooting
DRYSTAR 5500/5503 REFERENCE MANUAL
Overview of calibration functions
Three calibration functions are provided for the Drystar 5500.
Understanding the calibration policy and the calibration initiation
You should regularly perform a film calibration to assure that the image
quality remains optimal.
Film calibration can be initiated in 2 ways:
Automatically, i.e. after loading every ‘x’ new film packs (automatic
initiation),
Manually, i.e. you decide for yourself when to perform a film calibration,
e.g. daily, once a week, or only when the image quality tends to degrade.
For setting automatic or manual film calibration initiation, refer to ‘Changing
general image quality settings - Film calibration’ on page 91.
Calibration Purpose When?
1 ‘Film calibration’
To measure and set the
system density of the
printer.
To measure and set the
film sensitometry values.
•Automatic.
After an explicit
action of the
operator.
2 ‘Print head cleaning’
To eliminate white dots or
lines in transport direction.
•When image
quality problems
occur.
3
‘Print head profile
calibration’
To reduce vertical density
irregularities on the film.
After an explicit
action of the
operator.
Some calibration functions are interrelated, so do not perform any of the
above calibration functions as an isolated procedure.
Before performing a calibration, read the general guidelines as described in
‘Understanding the calibration policy and the calibration initiation’ on page 221.