User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
195
2900H EN 20071108 Troubleshooting
DRYSTAR 5500/5503 REFERENCE MANUAL
Checking the status indicator LED
The LED on the front panel indicates the Drystar 5500 operating status. Use
the following table to determine the status of your printer.
Color Light Status
Receive
jobs?
Printing
OK?
Action
Green
Constant
Ready (standby) Yes Yes Proceed.
Blinking
Start-up
sequence
No No
Wait.
Receiving jobs Yes Yes
Calculating Yes Yes
Printing Yes Yes
Calibrating Yes No
Installation No No
Proceed with
the activity.
Key-operator Yes/No Yes/No
Red
Blinking
Service mode No No
Upper/Lower
Input tray empty
Yes Yes/No
Refer to
‘Loading films’
on page 46.
Film jam Yes No
Refer to
‘Clearing of film
jams’ on
page 199.
Warning Yes Yes
Refer to
‘Warning
messages’ on
page 225
Constant
Error status No No
Refer to
‘Checking error
messages’ on
page 196.