User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
160
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Verifying Acceptable Spatial Resolution and Artifact Levels and Low Con-
trast Visibility for mammography application (DT 2 Mammo) (optional)
To verify acceptable spatial resolution, artifact levels and low contrast
visibility, proceed as follows:
1 Print the QC mammography test image or use the previously printed QC
mammography test image used to establish the daily operating density levels.
2 Visually check the test image for artifacts: no significant disturbing artifacts
should be visible.
3 Check the spatial resolution in each of the three ovals. Within each oval there
are three groups, each having five dots. All five dots of each group must be
visible with a magnifying glass. The smallest cluster of 5 dots is only visible if
the viewing conditions are good.
4 Check the Low Contrast Visibility at both the high (100 / 95%) and low end
(0 / 5%) of the density scale. You should be able to see the circle in the square
(refer to item 1 on the ‘QC mammography test image’ on page 158) and the
upper circle (refer to item 2 on the ‘QC mammography test image’ on page 158).
5 Record these values at the top of Chart 3 (‘Drystar 5500/5503 Artifacts and
Spatial Resolution Control Chart’). Refer to ‘Charts for mammography QC
(optional)’ on page 247.
6 These charts will be used for the weekly quality test. For more information,
refer to ‘Performing the Weekly QC tests for mammography application
(DT 2 Mammo) (optional)’ on page 163.
Good viewing conditions are important for the correct interpretation of
both diagnostic and test images. Make sure that the light box intensity
(luminance) is between 3000 and 6000 cd/m² (4500 and 6500 °K) for
mammography. Use a magnifying glass and use shutters to collimate.
Make sure the ambient light is low.
In case of significant artifacts or insufficient spatial resolution, the cause
must be found and the problem solved before any further clinical films can
be printed.
Refer to ‘Maintaining image quality and resolving Image quality problems’ on
page 220 and ‘Preventive maintenance schedule’ on page 186, or call your
local Agfa service organization.