User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
135
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Entering modality-specific settings
An instruction screen shows the information you need to complete this part of
the installation:
1 Press the Confirm key to continue or press the Escape key to skip this step.
• When you have pressed the Confirm key, the ‘Enter daily modality name’ screen
will appear:
2 Enter the modality nickname by means of the arrow keys. Refer to ‘Data entry’
on page 33. Make sure not to enter more than 8 characters.
3 Press the Confirm key to continue, or the Escape key to return to the previous
step.
The ‘Select modality brand name’ screen is displayed:
4 Press the Up/Down arrow keys to select the appropriate modality brand name.
5 Press the Confirm key to continue, or the Escape key to return to the previous
step.
The ‘Select modality type’ screen will be displayed:
2 Acuson
2 Philips
3 Trex
4 GE
1 MXVIEW
2 MX8000