User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
128
2900H EN 20071108Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Starting the Installation wizard
1 Press the Key-operator key to enter the Key-operator mode.
2 On the Key-operator main menu, press the Down key seven times, followed by
the Confirm key to select ‘Installation’.
The ‘Installation’ screen will be displayed:
3 Press the Down key once, followed by the Confirm key to select ‘Printer instal.
wizard’.
The ‘Select language’ screen is displayed:
4 Press the Up/Down arrow keys to select the desired language, followed by the
Confirm key.
The ‘Set date and time’ screen is displayed:
The number of displayed languages depends on the installed dataset. Contact your
local Agfa service organization for the latest Drystar 5500 language availability
status.
7 Installation
8 Quality control
1*English
2 Nederlands
3 Francais
4 Deutsch