User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
121
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
Installation
Upgrading the software
It is possible to install new software or new software components from a
floppy disk onto the hard disk.
Different software (components) can be installed:
• language files,
• modality specific files,
• calibration files,
• ...
The new software will be activated after the Installation procedure. The
previously installed software components will be automatically removed
when installing new software or new data.
1 Press the Key-operator key to enter the Key-operator mode.
2 On the Key-operator main menu, press the Down key seven times, followed by
the Confirm key to select ‘Installation’.
The ‘Installation’ screen will be displayed:
3 Press the Confirm key to select ‘Install data-sets’.
The following screen is displayed:
Before the procedure starts, the system suggests you to update your back-up copy of
the current configuration. It is highly recommended to do this, so you can always
restore it in case something should go wrong.
If you do not wish to update your back-up copy, press the Escape key and go to
step 10.