User Manual

Table Of Contents
113
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
4 In case the film calibration has failed, the following message will appear.
Press the Confirm key to retry the calibration.
5 When the film calibration has completed succesfully, this indicates that the
calibration procedure has been executed succesfully. This does not always mean
that the result of the calibration is succesful.
Three cases are possible:
Calibration successful, results OK.
The obtained Dmax (x.xx) after calibration is displayed:
Calibration successful, but Dmax differs more than 2% from the target
value.
The obtained Dmax (x.xx) after calibration is displayed:
Calibration procedure succesful, but limited by film-Dmax. The obtained
Dmax does not meet the required system Dmax and is limited by the film
sensitometry.
The obtained Dmax (x.xx) after calibration is displayed:
6 Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise, the old calibration parameters
will be used.
Blinking
Blinking
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924