User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
113
2900H EN 20071108 Advanced operation (Key-operator mode)
DRYSTAR 5500/5503 REFERENCE MANUAL
4 In case the film calibration has failed, the following message will appear.
Press the Confirm key to retry the calibration.
5 When the film calibration has completed succesfully, this indicates that the
calibration procedure has been executed succesfully. This does not always mean
that the result of the calibration is succesful.
Three cases are possible:
• Calibration successful, results OK.
The obtained Dmax (x.xx) after calibration is displayed:
• Calibration successful, but Dmax differs more than 2% from the target
value.
The obtained Dmax (x.xx) after calibration is displayed:
• Calibration procedure succesful, but limited by film-Dmax. The obtained
Dmax does not meet the required system Dmax and is limited by the film
sensitometry.
The obtained Dmax (x.xx) after calibration is displayed:
6 Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise, the old calibration parameters
will be used.
Blinking
Blinking
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer
e.g. Macbeth TR924