User Manual
Table Of Contents
- Table of contents
- About this manual
- Introducing the Drystar 5500
- Basic operation (operator mode)
- Advanced operation (Key-operator mode)
- Overview of Key-operator functions
- Stopping the printing process
- Viewing printer information
- Changing the configuration settings
- Printing images
- Saving the configuration settings
- Restoring the configuration settings
- Performing the calibration procedures
- Installation
- Quality control for general radiography applications (DT 2 B & DT 2 C)
- Quality control for mammography application (DT 2 Mammo) (optional)
- Controlling the Drystar 5500 via a remote PC (with browser)
- System description
- Preventive maintenance and replacements
- Troubleshooting
- Equipment information sheet
- Drystar media density response data
- Quality Control Charts
- Remarks for HF-emission and immunity
- Remarks for HF-emission and immunity
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment.
- This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variabl...
- Remarks for HF-emission and immunity
- Index
11
2900H EN 20071108 Introducing the Drystar 5500
DRYSTAR 5500/5503 REFERENCE MANUAL
Q Number of output trays
The Drystar 5500 is delivered with 4 output trays, which can be assigned to
modalities in any combination.
Q A Quality Control software module is available for the Key-operator. The QC
procedure for general radiography applications has been designed to comply
with the grayscale reproduction constancy test, according to the international
standard IEC 1223-2-4. For more information, refer to ‘Quality control for gen-
eral radiography applications (DT 2 B & DT 2 C)’ on page 141. The QC proce-
dure for the optional mammography application has been designed to comply
with the Mammography Quality Standards Act (MQSA) of the FDA. For more
information, refer to ‘Quality control for mammography application
(DT 2 Mammo) (optional)’ on page 153.
Q Integrated A#sharp technology
A#sharp is a technology that enhances image sharpness for the Drystar 5500.
An A#sharp label on the upper tray shows that the imager has been upgraded
with this technology.
The A#sharp technology is present in Software version 3.00 and higher.
1 A#Sharp label
1