User Manual

,
© AGD Systems Ltd 2010. All rights reserved., the information contained in this document is the property of AGD Systems Ltd., and is supplied
without liability for errors or omissions. AGD315 User Manual Page 12 of 12
Figure 3 shows the process by which the user settings are configured:
Set ranging mode
Set 40 FPS
Set 22 degrees
(via LW port)
Enter serial
number
xxxxx-yyyyy
(via LW port)
Reboot unit
RS422 OK ?
YES
NO
START
Unit
calibrated ?
Calibrate unit
YES
NO
Values stored in
parameter
EEPROM
Connect to unit via
Livewire port
Values stored in
secure
EEPROM
Enable
RS422 port
Set baud rate
(115200)
Check
serial number
(via RS422 port)
FAIL (TEST 11)
xxxxx-yyyyy from
secure EEPROM
Serial
number OK
?
YES
NO
FAIL (TEST 12)
Enable ranging
diagnostic
message
Reboot unit
Message OK ?
YES
NO
FAIL (TEST 13)
END
TEST
11
TEST
12
TEST
13
Figure 3 Configuration Of User Settings