Specifications
45
[1] When your meter is deactivated, press the CHECK ( ) but-
ton once to enter menu mode. The word “MEM” is displayed.
[2] Press the DOWN (
) or UP ( ) arrow buttons
until the abbreviation “HYPO” is displayed.
[3] Press the CHECK (
) button to enter and edit
the Hypoglycaemic Warning Alarm settings.
[4] Set the Hypoglycaemic Alarm ON or OFF
with the DOWN (
) or UP ( ) arrow buttons.
Your selection will be highlighted by the blink-
ing of the symbol. Press the CHECK (
) button
to confirm your selection and move on to the
hypoglycaemic threshold setting.
[5] Set the hypoglycaemic threshold (flashing)
with the DOWN (
) or UP ( ) arrow buttons. Press the CHECK
(
) button to confirm your selection and return to the menu.
Setting Hypoglycaemic
Warning Alarms
15
CHAPTER
The Hypoglycaemic Warning Alarm alerts you when your glucose
test results are below the hypoglycaemic threshold you have set.
The meter is pre-set to a hypoglycaemic threshold of 3.9 mmol/L.
Always consult your healthcare professional for your recommend-
ed target ranges. When your glucose test results are below your
set hypoglycaemic threshold, the meter display will flash your
test results, the abbreviation “HYPO,” time, and date repeatedly.
WARNING: Your healthcare professional should recommend
what your hypoglycaemic limit is.