page 21 of 25 Report No. 55S072148/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
42 TABLE: normal temperature P
Supply voltage . : 90V, 264V Test Condition: Normal load
Ambient temperature . : 25.0 ºC,
24.7 ºC.
Measuring location
Measured temperature
[ºC] Remarks
(Limit)
Model: ER-192, with Power Adapter: Lien, LE-0312B130W; D/A inverter: TDK, P/N: TBD319LF-2;
LCD panel: NEC, P/N: NL128102BC29, Panel placed vertically
Test Voltage 90V 264V -
Measured
value at
ambient
[ºC]
Corrected
value at 40
ºC ambient
[ºC]
Measured
value at
ambient
[ºC]
Corrected
value at 40
ºC ambient
[ºC]
Ambient 25.0 40 24.7 40 -
T1 coil (Power Adapter) 67.6 82.6 79.7 95.0
95
T1 core (Power Adapter) 62.0 77.0 76.9 92.2
95
L901 coil (DC/DC Board) 60.2 75.2 60.5 75.8
105
L903 coil (DC/DC Board) 46.6 61.6 47.2 62.5
105
PWB near U16 (Main Board) 58.6 73.6 58.9 74.2
105
PWB near U45 (Main Board) 61.7 76.7 62.0 77.3
105
L2 coil (Main Board) 72.1 87.1 72.6 87.9
105
T1 coil (Inverter) 85.1 100.1 85.5 100.8
105
T1 core (Inverter) 71.9 86.9 72.3 87.6
105
Metal Enclosure outside 36.2 51.2 36.6 51.9
60
Supplementary information: The tests were conducted under VGA mode.
COR - indicates measurements taken using change-of-resistance method
44 TABLE: overflow, spillage, leakage, humidity, ingress of liquids, cleaning,
sterilization, disinfection
P
Test type and condition Part under test Remarks
To clean with the damp cloth Unit PASS
Humidity
30 ºC, 93%, 168 hrs Unit PASS
Supplementary information: